MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray… – Catheter center lumen was found to have an occlusion in… – Covidien LP

← Back to Medical Device Recalls

Recall Category: Medical Device Recalls

Recalling Firm: Covidien LP

Product: MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 8888102004HP

Reason for Recall: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Classification: Class I

Status: Ongoing

Report Date: 08/23/2023

Recall Initiation Date: 06/28/2023

Distribution: US Nationwide – Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom

Code Info: GTIN: 20884521128009 LOT Numbers: 2200400154

Product Quantity: 685 units

Recall Type: Voluntary: Firm initiated

Initial Firm Notification: Letter

Firm Location: Mansfield, MA, United States

Recall Number: Z-2372-2023

Event ID: 92660