COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic… – Due to specific lots of diluents contributing to elevated platelet… – Beckman Coulter, Inc.

← Back to Medical Device Recalls

Recall Category: Medical Device Recalls

Recalling Firm: Beckman Coulter, Inc.

Product: COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemaker Stainers. For use as a diluent in conjunction with DxH ECO Cell Lyse for counting and sizing blood cells. For use as a rinsing agent.

Reason for Recall: Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.

Classification: Class II

Status: Ongoing

Report Date: 02/11/2026

Recall Initiation Date: 11/10/2025

Distribution: Worldwide – U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Canada, Chile, Mexico, Saint Kitts and Nevis, Trinidad and Tobago, and Viet Nam.

Code Info: Catalog Numbers: COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 UDI-DI code: 15099590233686 Lot Numbers: 2510410 2510420 2510450 3557500 3557510 3557540 3557560 3557570 3557640 3557650 3557660 3557670 3557680 3557700 3557710 3557850 3557920 3557930 3558000 3558030 3558040 3558060 3558090 3558100 3558110 3558120 3558140 3558150 3558160 3558170

Product Quantity: 76,098 units (US-61843 and OUS-14255)

Recall Type: Voluntary: Firm initiated

Initial Firm Notification: Letter

Firm Location: Miami, FL, United States

Recall Number: Z-1221-2026

Event ID: 97992